Frequently asked questions

The zero count filter aka purge filter is used to two reasons.

  • To establish that your instrument can count zeros so you are assured that you are starting with a precision base line.
  • To purge clean a contaminated sensor.

TSI particle counters count the stated particle size and the between sizes up to the next larger size while in differential mode.  In Cumulative mode your instrument will count from the stated size along with all larger particles. 

Yes, use TrakPro softwaee and go to “setup my location” to name locations then upload location names to your Hand Held.

(Applies to all Portable & Handheld Particle Counters)

Connect the Hand Held to the PC with the supplied cable. Use TrakPro software to download the collected data from the Hand Held into the PC. After data is downloaded, it can be saved as a MS Excel file.

Often the selection of a particle counter for use in a cleanroom is done based upon the specifications and cost of the instrument.

Before getting into the details of the specifications it is important to look at what the instrument will be used for, the environments it will be used in, and who will be using the instrument. Without this information taken into consideration, a less then optimal choice of particle counter for the application could be made. Here are some items to consider prior to selecting a particle counter:

What type of environment will the particle counter be used in? Will it be used in an ISO Class 3 Cleanroom for routine particle counting or will it be used for verifying a flow bench is operating prior to a critical process?

What type of data is the particle counter expected to collect? Will this information be recorded as simple pass/fail or will the information have to be logged into a spreadsheet or database?

Will the operator be carrying the particle counter around and placing it on a critical work surface or will it be cart mounted?

Will this particle counter be used to certify cleanrooms and travel from location to location?

Will the particle counter be used to monitor the cleanroom on a continuous basis? Is the particle counter intended to interface with a Facility Monitoring System (FMS)?

Specifications:

Though all manufacturers use the same principle, the details of the design are what set one manufacturer apart from the rest. Things like sample flow rate, sensitivity, size range and number of counting channels, durability of the laser or laser diode, lifetime of the light source, the ability to hold calibration all are important factors to consider.

Sensitivity: The smallest size particle that can be detected.

Zero Count Level or False Count Rate: The number of falsely reported particles using filtered air at the optimum flow rate for a given amount of time. The correct reporting of this is number of particles per 5 minutes. (Expected Zero Count rate should be less then 1 count per 5 minutes)

Counting Efficiency: The ratio of the measured particle concentration to the true particle concentration. The true particle concentration is measured with a more sensitive instrument that has a counting efficiency of 100% at the minimum particle size of the instrument under test. A properly designed instrument should have a 50% counting efficiency.

Channels: This is the number of “bins” the particles are placed in based upon the respective size of each particle counted. Channels are represented in microns. For example, you may have a particle counter with 4 channels. This means that the particles can be counted and binned in 4 different channels. Examples of channels are: 0.1 µm , 0.2 µm , 0.3 µm, 0.5 µm , 1.0 µm , 5.0 µm .

Flow Rate: This is the amount of air that passes through the particle counter. This is typically represented in cubic feet per minute. Common flow rates are 1.0 cfm and 0.1cfm. The greater the flow rate, the larger the pump to pull the air and the bigger the particle counter.

All too often minimum size is chosen over the other criteria. Though this is an important consideration, other parameters should also be considered.

Typically the more sensitive instrument, the higher the initial investment, and the higher the maintenance cost. If the instrument is used in environments with extremely high concentration of particles, it may require frequent cleanings by service technicians.

By understanding the intended use of the particle counter and the specifications, a more educated decision can be made when selecting a particle counter.

 

A Building Management System (BMS) also known as a building automation system (BAS) is a computer-based control system installed in buildings that controls and monitors the building’s mechanical and electrical equipment such as ventilation, lighting, power systems, fire systems, and security systems.

The FMS provides a single point of configuration and data analysis. It offers the capability to view graphs, charts, maps, SPC, system status, sensor status, and more. Each of these can be customized and displayed simultaneously. The FMS provides superior data collection reliability by providing redundant data collection engines, redundant data storage and data viewing. The FMS system can monitor Particle, Temperature/Relative Humidity, Differential Pressure, CO2, and many other environmental sensors related to your cleanroom environment.

 

An aerosol particle counter works on the principal of either light scattering or light blocking. An aerosol stream is drawn through a chamber with a light source (either Laser Based Light or White Light). When a particle is illuminated by this light beam, it is redirected or absorbed. Light scattered by a single particle in a specific direction in relation to the original direction has a unique signature which relates to the size of the particle. This allows for sizing and counting of individual particles.

A particle counter is made up of 4 components:

  1. Light Source (Gas Based Laser, Solid State Laser Diode, High Intensity Light)
  2. Photo Detection Electronics
  3. Sample Flow System
  4. Counting Electronics

 

Why do I need to monitor the state of my cleanroom?

There are two main reasons why you would monitor your cleanroom. One is to verify that your process environment is performing at the required specifications. The other is to document historical data of the process environment. Cleanrooms are built to the specific requirements of the products to be built. Contamination has a negative effect on yield and quality of these products. The cleanroom environment is very dynamic with products and personnel moving and changing, thus contamination events can happen at any time. Monitoring will enable you to know when the environment is not safe for building the products at any time. The ability to recall historical data is a very valuable tool to have. Trend charts, and records of events can help define maintenance cycles of equipment and cleaning. In addition, post mortem analysis is simplified when referring to this data. These reports will serve as records that can be used to present supporting data to customers, government organizations, and regulatory committees.

Can’t I just hire some people to go around to each room once a day and measure the cleanliness of each area?

Historically most companies started out that way, but companies learned that determining the level of particles in a room is not as simple as taking one single reading. Particulate levels must be measured at multiple locations throughout the cleanroom wherever product is exposed. Cases where measurements have been taken near HEPA or ULPA filters may show low particle levels, and yet at the product level work surface particulate levels may be unacceptable, causing product defects and lowering yield. Cleanroom processes are dynamic. Particle shedding events are rarely predictable, and can happen at random times throughout the day. Monitoring this environment on a continuous automated basis will help to catch these types of issues before they get out of control or affect production. Millions of dollars worth of products can pass through a cleanroom on a daily basis depending on the type of facility. Finding out about a contamination problem a day or two late could be an expensive and/or catastrophic situation. In most cases, the Return of Investment (ROI) on the contamination monitoring equipment and the system can be achieved during the first few detected events.

How can I effectively monitor my cleanrooms?

First you have to identify what the risks are for your product and process. Also, you need to understand the design intention of the cleanroom. This is necessary, as you would want to choose the right instrument for the application. A class 10,000 (ISO Class 7) cleanroom would not be monitored effectively with a 0.1 micron. Using a 0.3 or 0.5 particle counter would be a better choice.

  • Determine critical locations (locations where contamination can have an effect on the product or a large number of products)
  • Determine busy locations (locations where the product is moving or being manufactured).
  • Make an assessment of the cleanroom particle data during the operational state. Collecting data in the operational state (when manufacturing is actually occurring) is important to determine the locations where you are at risk.
  • A contamination monitoring system will be able to track, record and alarm when out of control limits are reached and/or exceeded.

What size particles do I need to monitor for?

This is determined by your product or your process. Other factors such as coverage and budget come into play. The smallest size particle that could possibly affect your product or smaller is a good place to start. However budgets and coverage of the whole process must be taken into consideration as well. Keep an important factor in mind. Particle shedding events occur at more than one particle size. Having one particle sensor in a cleanroom with 0.1 micron or lower resolution vs. having ten particle sensors with 0.3 micron resolution in the same cleanroom may not be a better situation. You must look at what is going on and what the risks are. Sometimes, many less sensitive sensors are better than a few or one of the more sensitive sensors.

How do I determine the number of airborne particle sample points that I need to monitor in my cleanroom, and where should I locate them?

The location and number of monitoring points is primarily dictated by the requirements of the product and the production process as well as understanding how the system will be used. Please refer to the Tech Papers section on this website to download “Justifying a Continuous Contamination Monitoring System”.

How do I justify the expense of an Automated Monitoring System?

Please see our Tech Papers section on this website for “Justifying a Continuous Contamination Monitoring System.”

Why can’t I use my building automation or facility control system for monitoring my cleanroom?

Many customers of ours have asked this question, and from a technical standpoint, it is possible to do so. There are several reasons for not integrating solely to the building control system. The primary responsibility of a control system is to keep the facility running between a given set of parameters. Data inside the control limits is often not recorded. The facility group’s primary responsibility is to keep the facility running within the set-points, by maintaining and repairing utilities. Contamination monitoring is more of a primary function of a quality, engineering or contamination control group. Sensing points are placed with a different intent for the data.

Often we will integrate our LMS system to a building control system (BCS), in fact we encourage our customers to do so. Integrating this data (either uni-directional or bi-directional) to and from a BCS will increase the number of monitoring points on both systems giving the customer a better perspective of what is going on in the controlled environment.

What is the difference between a control system and monitoring system?

The main differences between these systems are the architecture and principles for which they serve. The building control system is designed and installed with the critical parameters of the facility in mind. The sensors for that system are installed and provide feedback allowing for automatic compensation for out-of-spec events. A monitoring system is geared toward monitoring the process and product level of information. The two systems may share some overlap in the general environment, but both are very necessary for tuning your process. Without the process level of monitoring, you will not be able to detect changes in process environment due to factors contributing from personnel, or equipment.

When or where should I use a discrete based particle counter system vs. a manifold type?

Both systems have their place and usefulness. A discrete based or remote particle counter system gives great resolution without missing any data. The cost per point is more expensive than a manifold system, but has a better chance to catch intermittent particle excursions. A manifold sampling system is much more cost efficient, but sacrifices resolution due to the time it takes to rotate one complete revolution on the manifold. The more manifold points you add to the manifold, the longer it takes to complete one cycle, and it is possible to miss events while sampling other locations.

When evaluating your process to determine whether a discrete sensor or manifold system should be used consider the following things:

  • How many points do you want to monitor?
  • What is your budget?
  • What is the longest duration that you can have without information at that location?
  • Is there a cycle time on the tool that you’re concerned about capturing? (Automation, process or cleaning cycles)
  • Is this a critical location in your process?
  • Are there considerable product volumes at this location?
  • Who do you have to report this data to?

In short, if the maximum time of one complete sampling cycle on your manifold exceeds your maximum duration you can have without data from a single point you should try to use a discrete sensor for that location. In many cases, Astech, FMS will recommend a hybrid of systems, or both a manifold and discrete system working together. The manifold system for non-critical, or lesser critical locations, and discrete sensors for more critical parts of the process. This will allow excellent coverage and data, and still be cost effective. 

 

People often think of contamination as particles, dust, dirt or some type of chemical residue. 

Examples could be:

  • Air or liquid borne particles
  • Gas phase or airborne molecular contamination
  • Electrostatic charge or electrostatic discharge
  • Electromagnetic interference
  • Fluctuations of cleanroom or minienvironment differential pressure
  • Fluctuations of cleanroom temperature & relative humidity
  • Vibration
  • Surface particles
  • Air changes

 